aa and ab-series analog sensors - 13 - www.nve.com phone: 952-829-9217 fax: 952-829-9189 quick reference: aa and ab-series for comparison and product selection purposes, th e following table lists all available aa and ab- series analog sensors, with so me of their key characteristics: magnetometers: part number linear range (|oe 1 |) sensitivity (mv/v-oe 1 ) maximum non- linearity (% uni. 2 ) maximum hyster- esis (% uni. 2 ) maximum operating temp (c) typical resis- tance (ohms) package min max min max aa002-02 1.5 10.5 3.0 4.2 2 4 125 5k soic8 aa003-02 2.0 14 2 3.2 2 4 125 5k soic8 aa004-00 5.0 35 0.9 1.3 2 4 125 5k msop8 AA004-02 5.0 35 0.9 1.3 2 4 125 5k soic8 aa005-02 10.0 70 0.45 0.65 2 4 125 5k soic8 aa006-00 5.0 35 0.9 1.3 2 4 125 30k msop8 aa006-02 5.0 35 0.9 1.3 2 4 125 30k soic8 aah002-02 0.6 3.0 11.0 18.0 6 15 150 2k soic8 aah004-00 1.5 7.5 3.2 4.8 4 15 150 2k msop8 aal002-02 1.5 10.5 3.0 4.2 2 2 150 5.5k soic8 gradiometers: part number linear range (|oe 1 |) resistor spacing (mm) maximum non- linearity (% uni. 2 ) maximum hyster- esis (% uni. 2 ) maximum operating temp (c) typical resis- tance (ohms) package min max ab001-02 20 200 0.5 2 4 125 2.5k soic8 ab001-00 20 200 0.5 2 4 125 2.5k msop8 abh001-00 5 40 0.5 4 15 150 1.2k msop8 notes : 1. oersted (oe) = 1 gauss in air. 2. unipolar operation means exposure to magnetic fields of one polarity, for example 0 to +30 gauss, or -2 to -50 gauss. bipolar operation (for example, -5 to +10 gauss) will increase nonlinearity and hysteresis
|